Analytical Method Development and Validation of Obeticholic Acid Along With Stability Indicating Method by Quality by Design Approach

Darshit Trivedi, Kavita Desai, Jigar Patel*

Department of chemistry, Faculty of Science, Gokul Global University, India.

Corresponding Author E-mail: jjjigar007@gmail.com

DOI : http://dx.doi.org/10.13005/ojc/380526

Article Publishing History
Article Received on : 29 Aug 2022
Article Accepted on : 10 Sep 2022
Article Published : 10 Oct 2022
Article Metrics
Article Review Details
Reviewed by: Dr. R. K Lal
Second Review by: Dr. Raja. T
Final Approval by: Dr. Bal Krishan Sharma
ABSTRACT:

A chromatographic method was developed to determine the of assay of Obeticholic Acid with the help of Quality by design approach. QBD is a regulated technique for the process development. In Quality by Design, The method development starts with the specific intentions. A simple method was developed focusing on two critical quality attributes for development of the sturdy and polished method. The implication of self-sufficient factors was resolved using Fisher’s statistical test for Analysis of the Variance (ANOVA) model that was predicted. The aim of the present work is to develop a simple, precise, accurate method for the estimation of obeticholic acid by reverse phase chromatographic technique. Improved chromatographic conditions were observed on Thermo C18 column (150mm x 4.6mm5µm), using the mixture of Orthophosphoric acid (0.1 % OPA) and Acetonitrile (ACN) in the ratio of 55:45% v/v as mobile phase at the flow rate of 1.0mL/min. The column temperature was set at 25°C and wavelength for detection of moiety was 210nm. The system suitability parameters of the instrument used were found within acceptance criteria as per the ICH guidelines. Linearity study was carried out over a range from 100 ppm to 300 ppm and R2 value was found to be 0.9995. The precision study was calculated by injecting six samples of same concentration and the % Relative standard deviation was observed to be 0.3%. The mean percentage recoveries for 50%, 100% and 150% concentration were found to be 100.5%, 99.9% and 100.8% respectively. The percentage purity of the commercial sample assayed using the proposed method was found to be 99.8%. The developed method was validated according to ICH guidelines - Q2 (R1) and all the results were present within the acceptance criteria, Hence the proposed method can be employed for regular quality control analysis of obeticholic acid in bulk and pharmaceutical dosage forms.

KEYWORDS:

Analytical Method development; Antipsychotic Drug; Chromatography; Design Expert; HPLC; Obeticholic Acid; QBD approach

Download this article as: 

Copy the following to cite this article:

Trivedi D, Desai K, Patel J. Analytical Method Development and Validation of Obeticholic Acid Along With Stability Indicating Method by Quality by Design Approach. Orient J Chem 2022;38(5).


Copy the following to cite this URL:

Trivedi D, Desai K, Patel J. Analytical Method Development and Validation of Obeticholic Acid Along With Stability Indicating Method by Quality by Design Approach. Orient J Chem 2022;38(5). Available from: https://bit.ly/3SRjc2j


 Introduction

Obeticholic Acid is a natural bile acid which was identified as most effective physiological ligand for farnesoid X receptor in 1999. Obeticholic acid got the approval from U.S. Food and drug Association for the treatment of primary biliary cholangitis in 2016. It got its approval as an orphan drug as it is rarely available and cannot be produced for profitable purpose without the government assistance. Obeticholic acid (OBE), also known as 6-ethyl-chenodeoxycholic acid, is a brand-new semi-synthetic bile acid analogue. OBE is the strongest FXR agonist, and the presence of an ethyl group in the 6-position makes it possible to increase the activity of FXR-dependent processes as therapeutic targets for human disease.1 Obeticholic acid is a dihydroxy-5 beta-cholanic acid, 3 alpha-hydroxy steroid, and 7 alpha-hydroxy steroid in its chemical form.2

Analytical method development and Analytical method validation are considered as fundamentals for each and every pharmaceutical development schedule. A chromatographic analytical method is developed to determine purity of the compound to be analysed. A successful method development make sure that the resources that are available in the laboratory are amended and the methods fulfil the manifesto that are mandatory at ascending stages of the method development of the drug. An appropriate definition of method validation can be penned as “The process of demonstrating that analytical procedures are suitable for their intended use.”3,4,5

A chromatographic analytical method development is considered to be an instinctive and tedious process. In this modern and competitive world, the companies hastens the drug development curriculum and the competent drug pass through this process and it is necessary that the method developed must be rapid and robust.Most of the analytical methods are developed manually following the OFAT (One Factor At a Time) process, where only one factor is taken in to the consideration and the further modifications are dependent on the results obtained from the variations made in the factor. This is the time consuming process and repeated till the any advancement is observed and other parameter can be used 6,7,8,9,10,11. The improvement can be made in this process for the reduction of time by applying QBD (Quality By Design) approach. QBD is applied using the software based DOE (Design Of Experiment) operations that depends on multivariate modelling to anticipate accordingly and produce the advanced analytical chromatographic method which are then performed on HPLC system, resulting in significant growth in productivity.12,13,14,15,16,17

Materials and Methods

Instrumentation and chromatographic conditions

HPLC system of Agilent technologies 1260 Infinity Quaternary LC system model attached to DAD detector, auto sampler, column oven and degasser was used during the method development, forced degradation and validation. Data were evaluated on Chromeleon software version 6.8. Thermo C-18 column (150mm*4.6mm*5µ) was used as stationary phase for separation. Ultrasonic bath, Sartorius analytical balance and Vacuum micro unit with 0.45µ membrane filter were used during the study.

Chemicals and Reagents

Obeticholic Acid API was received as a gift sample from reputed pharmaceutical company. HPLC grade Acetonitrile (Rankem), Orthophosphoric Acid (Rankem), Hydrochloric acid (Merck), sodium hydroxide (Rankem), Hydrogen peroxide (Rankem) was used during development and analysis.

Chromatographic Conditions

Separation was obtained in Thermo C-18 column (150mm*4.6mm*5µ) when column and sampler temperature were maintained at ambient temperature. The isocratic elution method was applied with Buffer: Acetonitrile (55:45)(%v/v) as a mobile phase.The flow rate was adjusted to 1 ml/min and the injection volume was 20µl. The maxima for Obeticholic acid was detected at 210nm. The peak purity of main peak was checked using DAD detector.

0.1% Orthophosphoric acid

1 ml of orthophosphoric acid was diluted in 1000ml milli Q water, mixed well and sonicated for 2 minutes.The Buffer was filtered through 0.45µ membrane filter.

Mobile phase preparation

Mobile phase was prepared by mixing 550 ml buffer and 450 ml acetonitrile.

Selection of Wavelength

The wavelength for detection of Obeticholic acid was selected by scanning 200ppm sample of Obeticholic Acid in diluent in UV spectrophotometer and the maxima for the drug was observed at 210nm.

Selection of Mobile Phase

The trials for the different mobile phases were injected in the HPLC system for finalizing of mobile phase. Mobile phases used were different proportion of water, Acetonitrile, Methanol and Buffers. It was concluded that mixture of 0.1% Orthophosphoric acid and acetonitrile in (55:45) v/v amount provided the gratifying results which passes the criteria to be followed according to ICH guidelines. Thus the finalized mobile phase is 0.1% Orthophosphoric acid and acetonitrile (55:45 v/v).

Preparation of Standard Solution

Standard stock solution of 200ppm was prepared by dissolving 20mg Obeticholic Acid API in 100ml mobile phase. 20ppm diluted standard solution was prepared by diluting 5ml of standard stock solution to 50ml in mobile phase.

All the critical parameters used in the study passed the acceptance criteria according to ICH Guidelines. The system suitability criteria for the analytical method to be performed on HPLC are RSD%, tailing factor and theoretical plates. The solvent ratio and the flow rate were studied and optimized with design of experiments. The pH was not taken into the consideration as it did not show any major variation in retention time and the area of analyte during analysis. The effect of flow rate and the solvent ratio in the mobile phase in the variation of the results were studied. Most of the development studies follows OFAT (one Factor at a Time) but with the help of DOE multiple factors can be studied at one go. Thus the Central composite method, one of the widely used methods, is followed in QBD. Totally 13 experimental trials were implied by the software for analysing the cooperation of each level on formulation characters and the peak area (R1), tailing factor (5%) (R2) and theoretical plates were taken as dependent factors. The connotation of self-reliant factors was calculated using Fisher’s statistical test for Analysis of the Variance (ANOVA) model that was estimated.

Table 1: 22 Factorial design with high & low level factor statistical sequence optimization

Sr. No.

A:Mobile Phase (0.1% OPA) (ml)

B:Flow Rate ml/min

1

53

0.9

2

57

0.9

3

53

1.1

4

57

1.1

5

52

1

6

58

1

7

55

0.86

8

55

1.14

9

55

1

10

55

1

11

55

1

12

55

1

13

55

1

Validation Results

System suitability and precision

System suitability and precision were demonstrated by injecting six replicates of standard solution, prepared as per the test method. The peak area of analyte was recorded for replicate injections of standard solution.The precision was calculated by computing the relative standard deviation for the peak areaof replicate injections.

The observations are mentioned in Table 2.

Table 2: Replicates of Standard Solution.

Injection No.

Area of analyte

1

2908163

2

2899220

3

2908999

4

2889663

5

2898626

6

2909615

Average

2902381

% RSD

0.27

Table 3: System suitability observation

Parameters

Results

The relative standard deviation for peak area of Obeticholic Acidconcluded from six replicate injections of standard solution

0.27 %

Average tailing factor for Obeticholic Acid peak

1.17

Average theoretical plates for Obeticholic Acid peak

3950

 

The obtained results of the system suitability and precision are found within limit, which indicates that the system is correct and literal for analysis.

Method precision

Method precision was demonstrated by preparing six different samples of same batch as per the test method. The correctness of the method was studied by calculating the relative standard deviation for % Assay value. The results of the precision study are mentioned in table 4

Table 4: Method Precision Sample Replicates

Precision Sample Set  No.

% Assay

1

99.6

2

100.1

3

99.0

4

100.7

5

99.3

6

99.3

Average

100.3

% RSD

0.63 %

 

As the results shown are found to be within limit, thus it is concluded that method is suitable for determination of Assay in the sample.

Stability of analytical solution

Stability of analytical solutions (sample solution & standard) were established by preparing standard solution and sample as per method and by injecting at different intervals of time. The study was conducted at temperature 25°C up to 26 hours for standard solution and sample solution.

Table 5: Stability of standard solution at temperature: 25°C.

Time (hours)

Area

%RSD

Initial

2908163

NA

7

2899265

0.2

9

2894563

0.2

15

2896546

0.2

24

2885642

0.3

26

2889652

0.3

 

Table 6: Stability of sample solution at temperature: 25°C.

Time (hours)

Assay

Absolute Difference (%)

Initial

99.6

NA

7

99.4

0.1

9

99.8

0.2

15

99.6

0.2

24

100.6

0.5

26

99.0

0.5

 

Sample and standard solutionsare found stable up to 26 hours at 25°C temperature.

Linearity

The linearity of Obeticholic Acidwas carried out by preparing solutions of Obeticholic Acidstandard over the range of 50% to 150% of standard concentration. These solutions were analyzed into the HPLC system and area response of the solutions were reported. A graph of concentration Vs. Obeticholic Acidpeak area was plotted for the calculation ofthe correlation co-efficient between concentration and Obeticholic Acidpeak. The observations are mentioned in table 7.

Table 7: Linearity of Obeticholic Acid.

Linearity Level

Peak area

50%

1369141

80%

2254210

100%

2908135

120%

3578877

150%

4415850

Correlation coefficient (R)

0.9985

Resodual Sum Of Square (R2)

0.9970

The y-intercept (100% level)

-4.5278

Slope

14386.1700

Intercept

-61991.4000

 

Figure 1: A plot of concentration Vs. Obeticholic Acid peak area.

Click here to View figure 

The study shows that the response of Obeticholic Acidis linear from 50% to 150% of standard concentration.

Accuracy

The accuracy of the anlytical method was demonstrated by preparing recovery samples at the level of 50%, 100% and 150%. The recovery samples were prepared in set of three at each level. The above samples were analysed and the amountof API recovered from sample was calculated. The observations are tabulated in below Table 8,9 and 10.

Table 8: Recovery at 50% level

Sample No.

Amount spiked (mcg/ml)

Amount recovered (mcg/ml)

% Recovery

1

50

50.092

100.2

2

50

49.850

99.7

3

50

50.875

101.7

Average

100.5

%RSD

1.1%

 

Table 9: Recovery at 100% level

Sample No.

Amount spiked (mcg/ml)

Amount recovered (mcg/ml)

% Recovery

1

100

100.235

100.2

2

100

99.804

99.8

3

100

99.573

99.6

Average

99.9

%RSD

0.3 %

 

Table 10: Recovery at 150% level

Sample No.

Amount spiked (mcg/ml)

Amount recovered (mcg/ml)

% Recovery

1

150

152.146

101.4

2

150

149.961

100.0

3

150

151.595

101.1

Average

100.8

%RSD

0.8 %

 

As the recovery results obtained for Obeticholic Acidwas within the acceptable limits of recovery, it is been proved that the method is specific for quantification of Obeticholic Acidin the range of 50% to 150% of target concentration.

Specificity

The specificity of the test method was demonstrated by studying the interference from diluent and forced degradation samples. The samples with the stress conditions such as Acid degradation, Base degradation, Thermal degradation, Photolytic degradation and Oxidative degradation were prepared and injected. No degradation sample showed the interference with main peak.

Table 11: Degradation Data for specificity

Stress Type

% Assay

% Degradation

Peak Purity

As such

99.6

No degradation

Pass

Acid Hydrolysis

78.1

21.5%

Pass

Alkali Hydrolysis

92.2

7.4%

Pass

Peroxide Oxidation

97.7

1.9 %

Pass

Thermal Degradation

98.2

1.4 %

Pass

UV light Degradation

99.1

0.5 %

Pass

 

Robustness

The robustness was done by change in various parameters such as Flow rate of mobile phase, pH value of the buffer and the solvent ratio in the mobile phase. The samples were injected in triplicates. The results obtained are demonstrated in the following Table 11.

Table 12: Robustness data

Parameter

Condition

Mean Peak area

%RSD(%)

Flow rate

0.9 ml/min

2996604

0.46

1.1 ml/min

2734830

1.16

Solvent Ratio

47%

2865965

0.44

43%

3049977

1.38

Column Temperature

23

2789867

0.27

27

2856393

0.12

Figure 2: 3D presentation of for impact of Flow and OPA on Retention Time.

Click here to View figure 


Figure 3: 3D presentation of for impact of Flow and OPA on Theoretical Plates.

Click here to View figure 

Conclusion

A sturdy method for development of Obeticholic Acid API was developed on basis of Quality by Design approach on Design-Expert® Software, Version 12. Two self-reliant factors were taken in to consideration such as flow rate and solvent ratio. Total 13trial were recommended by the software for studying the synergy of individual level on the peak area, tailing factor and number of theoretical plate which the major response factors (dependent factors). The method validation was carried out following ICH guidelines. Specificity of the method was performed byanalysing degradation samples. The assay method for the API was prepared with present chromatographic condition developed, moreover it is found to be more authentic and decisive. The % Assay and % RSD appeared in range 100 ± 1.5% and <2, respectively, which meets the criteria mentioned in the ICH guideline. The stability studies were carried out and the samples are found to be stable for more than 26 hrs. According ICH guidelines, the forced degradation study was performed and no obstruction were observed because of degradation products elution time of the main product. Design Expert was able to automatically predict and test, optimized analytical methods.

Acknowledgement

With the deep sense of respect and gratitude, I express my indebtedness to my guide, Dr. Kavita Desai, Dr. Ritesh Vaidya Principal and Professor, Department of Chemistry and Dr. Jigar Patel, Gokul Science College, Gokul University Sidhpur for their valuable guidance, boundless enthusiasm and constant inspiration throughout the course of study.

Conflict of Interest

There is no conflicts of interest.

References

  1. Douša M, Slavíková M, Kubelka T, Černý J, Gibala P, Zezula J. HPLC/UV/MS method application for the separation of obeticholic acid and its related compounds in development process and quality control. Journal of Pharmaceutical and Biomedical Analysis. 149:214-24. 2018.
    CrossRef
  2. Kiranjyothi, R.; Chandrasekhar, K.B.; Balakrishnan, M.; Kiran, B.S., Stability Indicating RP-HPLC Method Development and Validation for the Quantification of Obeticholic Acid in Bulk and its Pharmaceutical Dosage Form. Indian journal of pharmaceutical education and research. 1; 54(2):S344-9, 2020
    CrossRef
  3. Singh, R., Journal of Pharmaceutical Education Research, 4(1), 20-27, 2013
    CrossRef
  4. Avellant, J. why Quality by Design an Executive’s Guide to the FDA’s Quality by Design G, Cerulean Associates LLC, 2008.
  5. Berridge, J. Q8(R1) Pharmaceutical Development Revision 1, CDER Reviewer Guidance , Version 8.1. 2007.
  6. CDER, Reviewer Guidance, “Validation of Chromatographic Methods, 1994.
  7. Nash, R. A. An International third edition, 129, 135-145, 2004.
  8. Patil, S. International Journal of Pharmaceutical Quality Assurance, 4(2); 13-19, 2013.
  9. Roy, S., International Journal of Pharmaceutical and Biomedical Research, 3(2), 100-108, 2012
  10. Karmarkar S., Journal of Chromatographic Science, Vol. 49, 143-149. 2011
    CrossRef
  11. Maalankaet., Acta Poloniae Pharmaceutica and Drug Research, Vol. 66 (5), 461- 470, 2009
  12. Module 3 Quality 3.2.P.2 Pharmaceutical Development, Quality by Design for ANDAs: An Example for Immediate-Release Dosage Forms, Food and Drug Administration, April 2012.
  13. Mohammad Y. Research Journal of Pharmaceutical, Biological and Chemical Sciences, 1 (2), 121-128. 2000
  14. Patel, D., International Journal of Pharmaceutical Sciences Review and Research, 4(3), 154-162. 2010
  15. Rambabuet, C., Chemical Science Transactions, 2(1), 41-46. 2013
  16. Santhosam, S. D., Journal For Drugs and Medicines, 3(1), 29- 33. 2011
  17. Smith A, Der Pharma Chemica, 2(3), 309-315. 2010
  18. ICH Harmonized Triplicate Guideline: Validation of Analytical Procedures: Text and Methodology Q2 (R1), ICH Steering Committee, Step 4 of ICH process, 25-27 2005