Abstract
Comparison of Conventional HPLC with UPLC Method for Determination of Albuterol Sulfate and it’s Impurities in Pharmaceutical Formulation
P. N. Dalvi, Z. Filmwala, K. Avinash and T Bane*
Abstract:
A simple, rapid, gradient reversed phase Ultra Performance Liquid Chromatographic method with ultraviolet detection (UPLC-UV) was developed for analysis of Albuterol sulfate and its impurities in Albuterol sulfate inhalation solution pharmaceutical formulation. Chromatography was carried out on a Waters Acquity BEH C18 (1.7μm, 100mm.x 2.1mm) column using 0.02 M Potassium Dihydrogen Orthophosphate as mobile phase A and Acetonitrile as mobile phase B. at a flow rate of 0.150mL/min and a 210 nm detection. Results from table 3 and 4 showed that Ultra Performance Liquid Chromatography exhibited a rapid and separation efficiency superior to that of existing conventional HPLC method.
Keywords:UPLC; HPLC; Albuterol Sulfate; Time; Cost
Back to TOC