Abstract
Construction and Performance Characterization of Ion Selective Electrodes for Potentiometric Determination of Citapolarm Hydrobromide in Biological Fluids
Amani S . Alturiqi1,2, Samar O. Aljazzar1,2, Reda A. Ammar1,2 and Nuwair khalaf3
DOI : http://dx.doi.org/10.13005/ojc/330138
Abstract:
A coated-wire ion selective electrode based on ion-pair complex of citapolarmhydrobromide (CT)with sodium tetraphenyl borate as electroactive material in the presence of dioctylphthalate (DOP) as the plasticizing solvent mediatorwas prepared. The electrode displays Nernstian response of 56.83 mV/decade over the concentration range of 1.0×10-6 to 1.0×10-2mol/Lat 25 °C. The influence of membrane composition, pH of the test solutionand foreign ions on the electrode wereinvestigated. The electrode was successfully applied to determination of the drug in pure form, urine and serumby standard addition potentiometry.
Keywords:Citapolarmhydrobromide(CT); Ion selective electrodes; potentiometry
Back to TOC