An Internal Standard High-Performance Liquid Chromatography Method for Simultaneous Quantification of Candesartan and Hydrochlorothiazide in Combined Formulations

Faisal K. Algethami

Chemistry Department, College of Science, Imam Mohammad Ibn Saud Islamic University (IMSIU), Riyadh 11432, Saudi Arabia.

Corresponding Author E-mail: arupdattadb@gmail.com

DOI : http://dx.doi.org/10.13005/ojc/370509

Article Publishing History
Article Received on : 15-07-2021
Article Accepted on :
Article Published : 28 Sep 2021
Article Metrics
Article Review Details
Reviewed by: Dr. Shafiul Islam
Second Review by: Dr. Guna Sekar
Final Approval by: Dr. Attilio Naccarato
ABSTRACT:

The internal standard method is a versatile procedure that avoids misleading results caused by the instability of the chromatographic system or inexperienced workers. It is an effective way to judge the accuracy of any obtained data. As the detector responses of chlorzoxazone (CZN) resemble those of candesartan (CDZN) and hydrochlorothiazide (HCTZ), CZN was employed as an internal standard. Herein, a simple chromatographic method was established for quantification of CDZN and HCTZ. Isocratic elution was conducted using 1% premixed acetonitrile/formic acid (7:3 v/v) at a 0.8 mL/min flowrate. The separation of the three components was maintained using the universal 20 µL loop, and for further simplicity in application, the analysis was optimized at 25 °C. CDZN, HCTZ, and CZN were simultaneously monitored and quantified at 270 nm. The method developed here complies with all the validation limits according to the British Pharmacopoeia (BP), United States Pharmacopoeia (USP), and the guidelines of the International Council ForHarmonisation (ICH). The method proved to be linear in the range of 6.4 to 25.6 µg/mL and 5.0–20 µg/mL for CDZN and HCTZ, respectively, while the quantitation detection limits were less than 1.0µg/mL for both.

KEYWORDS:

Candesartan; Chlorzoxazone; Hydrochlorothiazide; HPLC; Internal standard; Isocratic

Download this article as: 

Copy the following to cite this article:

Algethami F. K. An Internal Standard High-Performance Liquid Chromatography Method for Simultaneous Quantification of Candesartan and Hydrochlorothiazide in Combined Formulations. Orient J Chem 2021;37(5).


Copy the following to cite this URL:

Algethami F. K. An Internal Standard High-Performance Liquid Chromatography Method for Simultaneous Quantification of Candesartan and Hydrochlorothiazide in Combined Formulations. Orient J Chem 2021;37(5). Available from: https://bit.ly/3obf68v


Introduction

Candesartan (1-hydroxyethyl 2-ethoxy-1-[p-(o-1H-tetrazole-5-yl phenyl)benzyl] -7-benzimidazole carboxylate, cyclohexyl carbonate) (CDZN) and hydrochlorothiazide (6-chloro-3,4-dihydro-2H-1,2,4-benzothiadiazine-7-sulphonamide 1,1-dioxide) (HCTZ) are diuretics and angiotensin II receptor blockers prescribed for hypertension (Figure 1)1–3. Currently, researchers are interested in developing assay methods to validate their effectiveness, and though several have been developed for HCTZ and CDZN, most tests for these compounds separately, and only a few examine both substances simultaneously.Some researchers use spectrophotometric methods because of their ease of use 4,5, whereas others use LC–MS/MS, which would be ideal if not for the high cost of the equipment 3. High-performance liquid chromatography–ultraviolet (HPLC–UV) can be used in place of spectrophotometric procedures as it includes separation, and it is more affordable than the LC–MS/MS. Based on these reasons, an increasing volume of work has explored the use of HPLC–UV in assaying pharmaceutical formulations 6–11.

Method validation is a valuable topic in quantitative analysis because it is essential for demonstrating the reliability of any innovative analytical methodology 12,13. Because of the importance of this process, many agencies have set up a standard procedure 14–16. Method validation is an exclusive protocol that tests the accuracy, specificity, precision, reproducibility, limit of detection (LOD), and limit of quantification (LOQ) of a method to establish its suitability for a given purpose 17. The internal standard method is a powerful technique avoids the generation of misleading results caused by the detector or separation system 18.

This research was aimed at developing an HPLC photodiode array (HPLC–PDA) for simultaneously quantifying CDZN and HCTZ in pharmaceutical formulations. The muscle relaxant chlorozoxazone (5-chloro-3H-1,3-benzoxazole-2-one) 19 (CZN) presented in Figure 1c was used as an internal standard substrate to ensure the production of reliable results. As a result, this study presents a simple and fast analytical approach for application in quality control and research analysis.

Figure 1: The chemical structure of (a) candezartan, (b) hydrochlorothiazide, and (c) chlorozoxazone.

Click here to View figure 

Experimental

Chemicals

Working standards of CDZN, HCTZ, and CZN active pharmaceutical ingredientswere supplied from Ranbaxy Laboratories Limited, India. The combined tablets containing 16 mg CDZN and 12.5 mg HCTZ were purchased from a local market in Riyadh, KSA. CZN is an analgesic that was employed as an internal standard. HPLC-grade solvents acetonitrile was purchased from Sharlau, Spain. The placebo of the drug formulation was purchased from the Tabuk factory for pharmaceutical industries in Riyadh.

Preparation of Solutions

For preparation of the mobile phase solution, 700 mL acetonitrile and 300 mL formic acid were combined, and the solution was vacuum filtered, sonicated for 30 min, and cooled to 25 °C. CDZN (0.16 g) and HCTZ (0.125 g) were transferred to a 100 mL volumetric flask, which was half-filled with the mobile phase solution, sonicated for 30 min, and cooled to 25 °C. It was then filled to the mark with the same solvent. The CZN (300 mg/L) to be utilized as an internal standard substrate was prepared in the same manner. The appropriate stock solution volume was mixed with an aliquot of CZN stock solution, then diluted to 100 mL to attain a concentration of 16 µg/mL CDZN, 12.5 µg/mL HCTZ, and 30 µg/mL CZN. The average weight of 20 tablets (0.2596 g) underwent similar procedures to prepare a stock. Subsequent dilution was carried out, and the appropriate volume of CZN was added prior to the solution being adjusted to the appropriate final volume. Placebo pow-der (0.2311 g) was transferred to a 100 mL volumetric flask and processed similarly for sample preparation. A 0.22 µm nylon filter, which does not affect the concentration of solutions, was employed to filter solutions prior to their injection into the Shimadzu autosampler.

Chromatographic Conditions and Method Validation

The validation was conducted utilizing an HPLC/UV–Vis (Prominence, Shimadzu, Japan). Several columns were tested, including phenyl-hexyl, octyl (C8), octadecyl (C18). To determine the best elution, the mobile phase composition was varied in terms of the formic acid/acetonitrile ratio. Furthermore, column temperatures ranging 20–40 °C were tested,
as were flowrates ranging 0.5–2.0 mL/min. CDZN, HCTZ, and CZN were moni-tored at 270 nm.

The previously described standard solution was employed as a test for system suitability. The linearity and accuracy tests were performed using serial concentrations in the range of 40–160% of the target concentration. Typically, CDZN concentrations were 6.4, 9.6, 12.8, 16, 19.2, 22.4, and 25.6 µg/mL, while 5.0, 7.5, 10, 12.5, 15, 17.5, and 20 µg/mL concentrations were used for HCTZ. The solutions for linearity tests were prepared as a mixed standard, and each was injected with an internal standard aliquot to a final concentration of 30 µg/mL CZN. The detection and quantification limits were calculated statistically from the linearity results [16]. The standard, sample, and placebo solutions at optimized chromatographic conditions were used to investigate the procedure selectivity. Synthetic samples with 80 and 120% of the target concentrations were employed to test for robustness through intraday and interday studies. The three solutions were analyzed five times on the same day and on three consecutive days for the interday determination. The method robustness was tested by examining the effects resulting from slight alterations to the optimized chromatographic conditions. The detection wavelength varied by ±5 nm; the mobile phase composition was altered by ±5%, and the column temperature was varied by ±5 °C. Each parameter was changed separately, the analysis was conducted, and the recovery percentage then calculated.

Results

The separation of CDZN, HCTZ, and the internal standard CZN was accomplished using a mobile phase consisting of acetonitrile/formic acid (7:3 v/v) with a pH of 2.8. A flowrate of 0.8 mL/min was the best within the tested range, and the phenyl-hexyl column performed better than C8 and C18 in separating this combination. Furthermore, the PDA detector revealed that the best response for the three components was at 270 nm. Moreover, the commonly used 20 µL loop volume was selected to ease the procedure’s applicability.

To validate this method, the USP protocol was followed. A system suitability test that used a mixed standard solution of the target concentration was conducted. As monitored in Table 1, the statistical parameters of CDZN and HCTZ were within the acceptance criteria, implying the method’s repeatability.

Table 1: Statistical results for the system suitability investigation.

 

HCTZ

CDZN

CZN

HCTZ (corrected)

CANDI(corrected)

STD 1

1,284,626

146,064

2,947,909

0.435775

0.049548

STD 2

1,274,131

146,875

2,948,752

0.432092

0.049809

STD 3

1,279,409

147,677

2,948,303

0.433948

0.050089

STD 4

1,277,206

147,013

2,948,774

0.433131

0.049856

STD 5

1,279,045

147,166

2,948,493

0.433796

0.049912

STD 6

1,278,067

148,323

2,948,280

0.433496

0.050308

Avg.

1,278,747

147,186.3

2,948,419

0.433706

0.04992

STDV

3441.925

764.1706

327.1897

0.001211

0.000258

RSD

0.269164

0.519186

0.011097

0.279161

0.517656

Avg. tailing factor

1.425

1.658

1.367

 

Avg. resolution

8.936

4.847

4.947

Avg. theoretical plates number

5957.6

8014.6

10,777.8

 

The area of the analytes was divided by that of the internal standard to be corrected. Figure 2 illustrates the regression lines driven from the corrected peak area versus the concentration plot. The results show that the method was linear within the examined concentrations, and the correlation coefficient for both pharmaceutical ingredients was 0.999.

Detection and quantitation limits were determined and considered as an integral part of the validation protocol. The LOD and LOQ were statistically calculated using the results of linear regressions via Equations (1) and (2), respectively.

Figure 2: The linearity relationships of the concentrations and their corrected area for
(a) hydrochlorothiazide and (b) candesartan.

Click here to View figure 

The root mean square error (RMSE) was driven using the LINEST function (Microsoft Excel 2019): 0.0064 and 0.0006 for CDZN and HCTZ, respectively. For CDZN, the LOD was found to be 1.9 × 10−2 µg/mL, while the LOQ was 6.4 × 10−2 µg/mL. On the other hand, the values for HCTZ were 0.2 × 10−2 and 0.6 × 10−2 µg/mL, respectively.

The selectivity of the procedure was examined using the standard, sample, and placebo solutions in which the sample and standard solutions were injected with the internal standard. The results are shown in Figure 3. Figure 3b,c show that the excipients caused no alteration to the retention time of the active ingredients or the internal standard. Figure 3a shows two peaks with a negligible area at the retention time of HCTZ and CZN. These peaks could be attributed to cross-contamination from samples or standard injections. These results were considered acceptable since the areas were insignificant compared to the sample peak areas 14,16.

Further more, the accuracy of the method was investigated, revealing excellent consistency in the recovery values for a concentration range of 80–120% for both drugs. The average recovery for HCTZ and CDZN was 99.6 and 100.1%, respectively. The RSD values for these recoveries ranged between 0.8 and 1.6, showing that the results for both drugs were within the acceptance criteria (RSD ≤ 2%). These findings were comparable to, or better than, some results found in the literature for earlier methods 18-20.

Figure 3: Selectivity test using the placebo, the standard solution at target concentration, and the sample at target concentration injected with internal standard.

Click here to View figure

Moreover, the precision study was performed using the solutions prepared for the accuracy test. The repeatability of the internal standard method was inspected by conducting five consecutive assays for the test solutions. The procedure showed assay results in the range of 98.7–101.5 and 98.2–100.7 for HCTZ and CDZN, respectively. It is worth mentioning that the RSD values of the intraday investigations were within the acceptance criteria for both drugs and that the reproducibility of the procedure was tested by conducting the assay test on three different days. The results consistently reflected the method’s reliability (Table 2) 16,20.

Table 2: Precisions results for hydrochlorothiazide and candesartan using chlorzoxazone as an internal standard.

HCTZ

CDZN

Repeatability

Assay No.

80%

100%

120%

80%

100%

120%

1

101.5

100.4

100.8

98.9

99.4

99

2

98.7

99.3

100.2

98.4

99.1

99.8

3

101.6

100.6

99.4

98.2

100.6

99

4

101.5

99.2

99.1

100.2

98.4

98.2

5

101.7

99.8

99.6

99.5

100.7

99.1

Average

101

99.86

99.82

99.04

99.64

99.02

%RSD

1.3

0.6

0.7

0.8

1.0

0.6

Reproducibility

1

100.8

100

100.8

99

99.1

100.0

2

100.4

101

100.4

100

101.0

101.0

3

100.9

100

100.2

99

100.0

99.0

Average

100.7

100.3

100.5

99.3

100.0

100.0

%RSD

0.3

0.6

0.3

0.6

1.0

1.0

The robustness of the internal standard method was investigated by changing various parameters of the optimized conditions separately. The method performance was relatively consistent even during the slight changes of the chromatographic parameters, as shown in Table 3. The RSD value for the assay at the optimized and altered parameters was less than 2%. In addition, the average assay results of 99.7 and 100.8 for HCTZ and CDZN, respectively, demonstrate the robustness of this method.

Table 3: Test of method robustness based on measurement of HCTZ and CDZN using the developed internal standard method under optimized and altered chromatographic conditions.

No

CONDITION

HCTZ

Recovery %

CDZN

Recovery %

1

Optimized conditions

99.3

100.2

2

Mor 5 degree   Celsius

100.2

101.9

3

less 5 degree   Celsius

100.1

100.5

4

5% More flow rate

99.4

99.7

5

5% less flow rate

99.0

102.9

6

5% more Organic solvent

100.2

102.0

7

5% less Organic solvent

99.5

99.3

8

More 3 nm

99.8

100.4

9

Less 3 nm

99.8

100.5

Avg

99.7

100.8

% RSD

0.428

1.2

Assay of local-market tablets by the developed method

The labeled content of HCTZ and CDZN in the tablet was 16 and 12.5 mg, respectively. The sample solution was injected with the internal standard (CZN), completed to 50 mL with the same solvent, and analyzed at optimized chromatographic conditions. The content of the HCTZ and CDZN tablets was determined to be within the acceptable range with an assay of 100.2 and 103.7%, respectively.

Conclusions

A reversed-phase HPLC–PDA procedure was developed for the simultaneous quantification of HCTZ and CDZN combined in pharmaceutical formulations in which CZN was employed as an internal standard. Compared to conventional HPLC methods, the method developed here has the advantage of allowing self-correction via inclusion of the internal standard. The validated internal standard method successfully bypassed all the obstacles associated with fluctuations in the detector, temperature, pump flow, and mobile phase composition, thereby showing great potential for application in routine analysis and research.

Conflict of Interest

The author declares that he has no conflict of interest.

References

  1. Khedr, A.E. Simultaneous Determination of Candesartan Cilexetil and Hydrochlorothiazide by High-Performance Liquid Chromatography. J. King Abdulaziz Univ. Med Sci. 2008, 15, 3–13.
    CrossRef
  2. Srikalyani, V.; Madhuri, T.; Sareesh, K.; Kumar, R.M. Simultaneous estimation of hydrochlorthiazide and candesartan cilextil in bulk and pharmaceutical dosage forms by RP-HPLC PDA method. Int. J. Pharm. Sci. Res. 2018, 17, 150–157.
  3. Stolarczyk, M.; Maślanka, A.; Apola, A.; Kwiecień, A.; Hubicka, U. Determination of candesartan, hydrochlorothiazide and rosuvastatin s active ingredients of a polytablet by a chromatographic-densitometric method. Acta Pol. Pharm. 2020, 77, 833–840.
    CrossRef
  4. Marghany, K.A.; Salamorcid, R.A.; Hadadorcid, G.M.; Ibrahim, E. Green UV Absorbance Ratio Method for Determination of Eprosartan and Hydrochlorothiazide. Rec. Pharm. Biomed. Sci. 2021, 5, 90–99.
    CrossRef
  5. Grewal, I.K.; Singh, S.; Arora, S.; Sharma, N. Development and validation of UV-spectrophotometer analytical method of eflornithine hydrochloride. Plant Arch. 2020, 20, 3265–3270.
  6. el Karbane, M.; Benchekroun, Y.H.; Abousalih, F.Z.; Bennani, I.; Azougagh, M.; Saffaj, T.; Bouatia, M. Development and val-idation of a UPLC-DAD method for the simultaneous quantification of eight antihypertensive drugs in the pharmaceutical matrix. J. AOAC Int. 2020, 104, 562–570.
    CrossRef
  7. Patel, M.; Kothari, C. Quantitative bio-analysis of pitavastatin and candesartan in rat plasma by HPLC-UV: Assessment of pharmacokinetic drug-drug interaction. J. Chromatogr. B 2020, 1138, 121962.
    CrossRef
  8. Tan, A.; Gui, X.; Wong, M.; Deng, H.; Gu, G.; Fanaras, C.; Fanaras, J.C. Simultaneous quantification of candesartan and irbesartan in rabbit eye tissues by liquid chromatography–tandem mass spectrometry. Biomed.Chromatogr.2020, 34, e4808.
    CrossRef
  9. Bulk, A.I. Int. J. Mod. Pharm. Research.
  10. Marghany, K.A.; Abdelsalam, R.A.; Haddad, G.M. HPLC method transfer study for simultaneous determination of seven angiotensin II receptor blockers. J. Sep. Sci. 2020, 43, 1398–1405.
    CrossRef
  11. Janet, B.S.; Suthakaran, R.; Kymar, B.P.; Chankar, C.H.; Syed, G.; Suneetha, K. A Prospective Validation of Hydrocholoro-thiazide and BisoprololFumerate in its Pure and Bulk Dosage Forms by using RP-HPLC Technique. Asian J. Pharm. Anal. 2020, 10, 1–6.
    CrossRef
  12. Hopfgartner, G. Bioanalytical Method Validation: How Much Should We Do and How Should We Document?; Springer: Ber-lin/Heidelberg, Germany, 2020.
    CrossRef
  13. Ozkan, S.A. Analytical method validation: The importance for pharmaceutical analysis. Pharm. Sci. 2018, 24, 1–2.
    CrossRef
  14. Q2 (R1) validation of analytical procedures. In ICH Quality Guidelines; Teasdale, A., Elder, D., Nims, R.W., Eds.; Wiley: Hoboken, NY, USA, 2017; pp. 127–166.
    CrossRef
  15. Pharmacopeia. General Chapters:< 1225> Validation of Compendial Procedures; 2011; pp. 778–782.
  16. ICH. Q2 (R1): Validation of analytical procedures: Text and methodology. In International Conference on Harmonization; ICH: Geneva, Switzerland, 2005.
  17. Raposo, F.; Ibelli-Bianco, C. Performance parameters for analytical method validation: Controversies and discrepancies among numerous guidelines. TrAC Trends Anal. Chem. 2020, 129, 115913.
    CrossRef
  18. Gülfen, M.; Canbaz, Y.; Özdemir, A. Simultaneous Determination of Amoxicillin, Lansoprazole, and Levofloxacin in Phar-maceuticals by HPLC with UV–Vis Detector. J. Anal. Test. 2020, 4, 45–53.
    CrossRef
  19. Roy, P.; Ghosh, A.J.C. Mechanochemicalcocrystallization to improve the physicochemical properties of chlorzoxazone. CrystEngComm 2020, 22, 4611–4620.
    CrossRef
  20. Puumalainen, I.; Halonen, P.; Johnson, K.; Enlund, H.; Marja Validation of the United States Pharmacopeia (USP) medication counselling behaviour guidelines. Airaksinen Pharm. Educ. 2005, 5, 87–96.
    CrossRef

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

About The Author