Simple, Sensitive, and Easy to use Method for Determination of Palladium Content in Palladium Acetate Using Flame Atomic Absorption Spectrophotometry
Melwin Diego D’souza1*, Poonam Dwivedi1, Rama Lokhande 1, Tushar Anvekar2 and Andrew Joseph D’souza2
1Department of Chemistry, Jaipur National University, Jaipur, 302017, India.
2Department of Chemistry, St Xaviers College, Mapusa -Goa-403507, India.
Corresponding Author E-mail: melwindsouza3890@gmail.com
DOI : http://dx.doi.org/10.13005/ojc/370415
Article Received on : 18-Jun-2021
Article Accepted on :
Article Published : 13 Aug 2021
Reviewed by: Dr. Qasim Mohmmed
Second Review by: Dr. Amira Salah
Final Approval by: Dr. Abdulwahab Omri
A simple, sensitive, and easy to use method for the determination of palladium content in palladium acetate was developed using homogeneous acid digestion of palladium acetate followed by atomic absorption spectrometry (AAS) analysis. Parameters affecting the extraction efficiency of palladium during digestion of palladium acetate were investigated and optimized. The palladium content was measured at 244.7 nm by atomic absorption (AA) spectrometers on PinAAcle™ 900 series with an air-acetylene flame. The linear calibration graph was observed in the range of 2.5-15 ppm with a limit of detection of 8.33 ppm and a correlation coefficient of 0.999970. For 5 replicate analyses of each palladium reference solution the relative standard deviation was within 2.00%.Analytical method validation for the method was carried out and obtained results were satisfactory.To verify, the accuracy parameter of the method validation, for the determination of Pd(II) from sample palladium acetate standard addition method was applied and satisfactory results were obtained. Optimization of analytical method variables during analytical method development is discussed.
KEYWORDS:Air-Acetylene Flame; Atomic Absorption Spectrometry; Palladium Acetate
Download this article as:Copy the following to cite this article: D’souza M. D, Dwivedi P, Lokhande R, Anvekar T, D’souza A. J. Simple, Sensitive, and Easy to use Method for Determination of Palladium Content in Palladium Acetate Using Flame Atomic Absorption Spectrophotometry. Orient J Chem 2021;37(4). |
Copy the following to cite this URL: D’souza M. D, Dwivedi P, Lokhande R, Anvekar T, D’souza A. J. Simple, Sensitive, and Easy to use Method for Determination of Palladium Content in Palladium Acetate Using Flame Atomic Absorption Spectrophotometry. Orient J Chem 2021;37(4). Available from: https://bit.ly/3CMPmnI |
Introduction
There is an increasing demand for palladium elements in today’s industries, and palladium is the widely used element of platinum group metals (PGM)1.Furthermore, palladium is used in various areas of science and technology e.g. metallurgy, and palladium is also used as a catalyst especially in dehydrogenation and hydrogenation of some of the important organic compounds2. High selectivity3-5, sensitivity, and the control effects of interference require analytical methods of palladium analysts6-8. Applications of palladium and its alloys, both in instrumentation and in the chemical industry were also studied before. A Chemical compound of palladium acetate of palladium described by the formula [Pd(O2CCH3)2]n. Determination of Pd content, spectrophotometry is one of the common methods, the main advantages of which is rapidity, accessibility, and simplicity. For the determination of palladium there exist many spectrophotometric methods, using inorganic or organic reagents, but they have some limitations. In environmental samples estimation of Pd contents, the most widely used methods include air flame atomic absorption spectrometry (AF-AAS)9,10 and inductively coupled plasma-mass spectrometry (ICP-MS)11,12. Spectrophotometric determination of palladium by the coloration with 2-mercaptoethanol reported13 earlier and Hussainien reported14 FAAS determination of palladium after its selective recovery by silica modified with hydrazones. With the availability of a variety of atomic spectroscopy methods, we developed a simple, sensitive, and easy to use method for the determination of palladium content in palladium acetate using flame atomic absorption taking into consideration a variety of important criteria, including sample throughput, interferences, detection limits, data quality, cost, analytical working range and ease of use. The results obtained indicate that the proposed method is highly sensitive, selective, and easy to use for the determination of palladium in palladium acetate samples.
Experimental
Chemicals and reagents
Diluent
5N Hydrochloric acid.
Dilute 425ml of concentrated Hydrochloric acid to 1000ml with water.
Blank solution
Diluent: 5N Hydrochloric acid used as blank.
Test solution
Weigh accurately about 100 mg of palladium acetate in a 250ml standard volumetric flask add 10 ml Conc. HCl and heat. Allow the flask to cool to room temperature and then make up the volume using 5 N Hydrochloric acid dilute upto the mark with diluent. Dilute 5.0ml of the solution to 100.0ml with 5 N Hydrochloric acids (about 10ppm of Palladium).
Reference stock solution
Dilute 10.0ml of 1000ppm (palladium standard for AAS trace CERT®, 1000 mg/L Pd in hydrochloric acid) of Palladium to 100.0ml with diluent (100ppm).
Reference solution
Prepare 5 different reference solutions ranging from 2.5ppm to 15ppm from reference stock solution as per the Table-1 given below,
Table-1: Preparation of reference solutions using a stock solution.
Sr.No. |
Reference solution in ppm |
The Volume of reference stock solution to be taken in ml |
Volume to be diluted with diluent in ml |
1 |
2.5 |
2.5 |
100 |
2 |
5.0 |
5.0 |
100 |
3 |
7.5 |
7.5 |
100 |
4 |
10.0 |
10.0 |
100 |
5 |
15.0 |
15.0 |
100 |
Instrumentation
For analysis, Perkin Elmer PinAAcle™ 900 series of atomic absorption (AA) spectrometer using air acetyleneflame. A palladium hollow cathode lamp with a 247.6nm wavelength, a 9.0mA current, and a0.4nm slit width was used as the radiation source.
Limit
The Palladium content in palladium acetate by analysis should be between (% w/w) 45.90 – 48.40%
Validation of the method
To ensure that the analytical methodology is valid and consistent for the determination of palladium content in palladium acetate by atomic absorption spectrometer (AAS). The method was validated various parameters such as LOD & LOQ, specificity, robustness, linearity, accuracy, and precision15. The above method has been fully developed and completely validated.
Results and Discussion
System Suitability
The replicate injections of standard solution were injected. The system suitability parameters are summarised in Table-3 and are well within the range of acceptance criteria (%RSD ≤ 5).
Table 2: Optimal operating conditions for flame atomization of palladium
Element |
Palladium |
Wavelength |
247.60nm |
Read Time |
10 sec |
Lamp Current |
9 mA |
Recommended Flame |
Air-Acetylene |
Fuel Gas Flow |
2.0 |
Support Gas Flow |
17.0 |
Slit Width |
0.4nm |
Signal Type |
Atomic Absorption |
Atomization Site |
Burner head |
Equation |
Linear Through Zero |
Flame Type |
Air-Acetylene |
Pre Spray Time |
3 sec |
Integration Time |
5 sec |
Table 3: Readings of Absorption for System Suitability.
Sr. No |
Pd Concentration mg/mL |
Absorption |
1 |
9.419 |
0.3092 |
2 |
9.343 |
0.3070 |
3 |
9.336 |
0.3068 |
4 |
9.168 |
0.3021 |
5 |
9.113 |
0.3005 |
6 |
9.158 |
0.3018 |
Average |
0.3046 |
|
SD |
0.0035 |
|
% RSD |
1.16 |
Specificity
In this specificity aspirated blank (5N Hydrochloric acid used as blank), standard solution (six replicates), test solution, and spiked test solution were used.The results recorded the absorbance and responses of all the above-mentioned solutions. The palladium was not included in the blank as a result the blank showed no absorbance at the selected wavelength of palladium hence no interference from the blank. The analyst checked the interference of blank with standard and the data is summarized in Table-4. There is no interference of palladium in blank solution and hence the proposed method is specific for the determination of palladium from palladium acetate16.
Table 4: Average concentration for Specificity test.
PalladiumConcentration (ppm) |
Wavelength |
Average concentration n=3 |
Blank Solution |
247.60 |
0.0007 |
Standard Solution |
247.60 |
0.3077 |
Sample Solution |
247.60 |
0.3015 |
Precision at LOQ Level
The following precision was reported as percentage relative standard deviation (%RSD) by aspirating the 8.33mg/L palladium standard 6 times for intraday and interday measurements. The results are presented in Table-5 and Table-6 respectively. The results were found to be within the range of acceptance criteria (%RSD ≤ 5)17. Hence method meets the requirements of System Precision.
Table 5:Readings of absorbance ofpalladium standard for precision at LOQ Level(Intraday).
Sr. No. |
Concentration (ppm) |
Absorbance |
1 |
8.33 |
0.2656 |
2 |
8.33 |
0.2655 |
3 |
8.33 |
0.2549 |
4 |
8.33 |
0.2555 |
5 |
8.33 |
0.2551 |
6 |
8.33 |
0.2656 |
Average |
0.2604 |
|
SD |
0.0057 |
|
% RSD |
2.19 |
Table 6: Readings of absorbance of palladium standard for precision at LOQ Level (Interday).
Sr. No. |
Concentration (ppm) |
Absorbance |
1 |
8.33 |
0.2475 |
2 |
8.33 |
0.2534 |
3 |
8.33 |
0.2612 |
4 |
8.33 |
0.2562 |
5 |
8.33 |
0.2483 |
6 |
8.33 |
0.2545 |
Average |
0.2535 |
|
SD |
0.0051 |
|
% RSD |
2.02 |
Linearity and Range
The analyst prepared 5 different concentrations of reference solution from palladium standard solution. The results were recorded and calculated the mean area for each solution. The analyst further calculated the slope, Y-intercept, and correlation coefficient. A graph was plotted showing the correlation of concentration against peak area. The data obtained are summarized in Table-7. The calibration curve is shown in Figure-1. The concentration of a standard solution is directly related and proportional to absorption in their lower and upper limits7-10. The calibration curve is shown in theregression equation. The ranges were plotted between palladium standard solutions 2.5-15.0ppm. The range plotted from LOQ level to 150% with respect to sample solution and was within range of acceptance criteria11-16 (Correlation Coefficient ≤ 0.99).The concentration range of palladium standard solutions was between 2.5-15.0ppm and the absorption results were recorded18.
Table 7: Absorption made by different concentrations of palladium standard solution
Concentration in ppm |
Absorption |
2.5 |
0.0916 |
5.0 |
0.1763 |
7.5 |
0.2519 |
10.0 |
0.3267 |
15.0 |
0.4534 |
Correlation coefficient |
0.99997 |
Figure 1: Calibration Curve of different concentrations of palladium standard solution. |
Accuracy/ Recovery
For the accuracy test, the palladium acetate sample was spiked with palladium standard at 3 levels each in triplicate, i.e. 3 x 150%, 3 x 100%, and 3 x 50%. Each sample was analyzed against astandard reference solution of palladium. From the amount added and the amount found, the percentage recovery was calculated in Table-8.The percent recovery of the added standard was found to be 98.34 %. The method is free from either positive or negative interferences from the blank. From the above result, it was found that the recovery data of the analyte is within the limit and hence the proposed method is accurate19.
Table 8: Percentage recovery for the accuracy test
Level (%)/ ppm |
Sample Wt (in mg) |
Conc (in ppm) |
Area (mv) |
% Recovery |
%Mean Recovery |
STD DEV |
% RSD |
50_1 |
5.3 |
5.30 |
0.1757 |
101.05 |
102.70 |
2.43 |
2.37 |
50_2 |
5.3 |
5.30 |
0.1766 |
101.56 |
|||
50_3 |
5.1 |
5.10 |
0.1765 |
105.49 |
|||
100_1 |
10.0 |
10.00 |
0.326 |
99.37 |
99.57 |
0.21 |
0.21 |
100_2 |
10.0 |
10.00 |
0.3266 |
99.55 |
|||
100_3 |
10.0 |
10.00 |
0.3274 |
99.79 |
|||
150_1 |
15.0 |
15.00 |
0.4537 |
92.19 |
92.75 |
0.64 |
0.69 |
150_2 |
14.9 |
14.90 |
0.4527 |
92.61 |
|||
150_3 |
14.8 |
14.80 |
0.4537 |
93.44 |
|||
Overall % Recovery |
98.34 |
||||||
Overall STDEV |
4.58 |
||||||
Overall % RSD |
4.66 |
Wt: Weight, Conc = Concentration, STD DEV = Standard Deviation, RSD = Relative Standard Deviation
Sample Analysis
The sample was prepared and aspirated as per the above method and the palladium content was found in the given sample is 46.43% which is within acceptance criteria20-21 (USP NF 40).
Conclusions
The above-validated method is specific for the determination of palladium content in palladium acetate. The method is linear, precise, and accurate across a suitable analytical range. The method was successfully applied to the determination of palladium content in palladium acetate samples and satisfied recoveries and suitable reproducibility were obtained. In this Flame AAS method, consumption of toxic solvents is minimum. The proposed method has reduced digestion time for the determination of palladium in palladium acetate samples22.
Acknowledgement
The authors are highly thankful to Perkin Elmer Mumbai, for allowing us to use their PinAAcle 900 series AA spectrometer.
Conflict of Interest
The authors declare that there is no conflict of interest regarding the publication of this paper titled Simple, sensitive, and easy to use method for determination of palladium content in palladium acetate using flame atomic absorption spectrophotometry.
References
- Greenwood, N. N. J. of the Chem. Soc, Dalton Trans.2001 (14), 2055–2066. doi:10.1039/b103917m .
CrossRef - Balcerzak, M. Analytical Sciences. 2002, 18(7), 737–750. doi:10.2116/analsci.18.737.
CrossRef - Bagheri, A.; Taghizadeh, M.; Behbahani, M.; Asgharinezhad, A. A.; Salarian, M.; Dehghani, A.; Ebrahimzadeh, H.; Amini, M. M. Talanta; 2012, 99, 132. https://doi.org/10.1016/j.talanta.2012.05.030
CrossRef - Salih, B.; Celikbicak, O.; Doker, S.; Dogan, M. Anal. Chim. Acta; 2007, 587, 272-278.https://doi.org/10.1016/j.aca.2007.01.063.
CrossRef - Zereini, F.; Alt, F. Palladium emission in the environment; Springer-Verlag: Berlin Heidelberg, 2006. DOI: 10.1007/3-540-29220-9_26
CrossRef - Sures, B.; Zimmermann, S.; Messershmidt, J.; Von Bohlen, A.; Alt, F. Environ. Pollut. 2001, 113, 341. https://doi.org/10.1016/S0269-7491(00)00185-8.
CrossRef - Sures, B.; Zimmermann, S.; Messerschmitt, J.; Von Bohlen, A. Ecotoxicology 2002, 11, 385. https://doi.org/10.1023/A:1020517624371.
CrossRef - Vanketel, W. G.; Ntebber, C. Contact Dermatitis 1981, 7, 331. https://doi.org/10.1111/j.1600-0536.1981.tb04092.x
CrossRef - Bencs, L.; Ravindra, K.; Van Grieken, R. Spectrochim. Acta, Part B. 2003, 58, 1723.
- Godlewska-Zylkiewicz, B. Microchim. Acta. 2004, 147, 189. available https://doi.org/10.1016/SO584-8547(03)00162-9.
- Gomez, M. B.; Gomez, M. M.; Palacios, M. A. Anal. Chim. Acta. 2000, 404, 285.https://doi.org/10.1016/S0003-2670(99)00723-0.
CrossRef - Rauch, S.; Motelica-Heino, M.; Morrison, G.M.; Donard, O. F. X. J. Anal. Spectrom; 2000, 15, 329. https://doi.org/10.1039/A909100I.
CrossRef - Nagaraj, P; Gopalakrishna, N; Chandrashekhara, K. Res. J. Chem. Sci. 2015, 5(9), 19-26. doi.org/10.5281/zenodo.3566353.
- Hassanien, M. M. Microchim Acta. 2009, 167,81. https://doi.org/10.1007/s00604-009-0219-2.
CrossRef - Mukherjee, P. K. In Quality Control and Evaluation of Herbal Drugs. 2019. DOI: 10.1016/B978-0-12-813374-3.00016-8.
CrossRef - Uesugi, K.; Sik, L. J.; Nishioka, H.; Kumagai, T.; Nagahiro, T. Microchem. J. 1994, 50(1), 88-93. https://doi.org/10.1006/mchj.1994.1062.
CrossRef - Shetty, N. A.; Gadag, R. V. Bull. Chem. Soc. Jpn. 1993 66(9), 2536- 2540. https://doi.org/10.1246/bcsj.66.2536.
CrossRef - Shetty, P.; Shetty, N. A.; Gadag, R. V. Indian J. Chem. Technol. 2003, 10, 287-290.
- Suvarapu, L. N.; Reddy, K. J.; Reddy, S. A.; Jyothi, R. K.; Reddy, A. V. J. Chin. Chem. Soc. 2007, 54(5), 1233-1241. https://doi.org/10. 1002/jccs.200700174.
CrossRef - Yadav, N.;Goyal, A. International Journal of Current Pharmaceutical Research., Dec. 2016, 9(1), 132-135
- Karthikeyan, J.; Parameshwara, P; Shetty, N. A. Environ. Monit. Assess. 2011, 173(1-4), 569-577. https://doi.org/10.1080/14756366.2016.1186020.
CrossRef - Reddy, J.K.;Rajesh, K. J.; Ramachandraiah, C.; Reddy, A. S.; Reddy, V. A. Environ Monit. Assess. 2008, 136(1-3), 337-346.
- ICH Guideline, Q2 (R1), Validation of Analytical Procedures: Text and Methodology, 2005.available.
- World Health Organization. Guidelines on validation – appendix 4 Analytical method validation, 2018. https://www.who.int/medicines/areas/quality_safety/quality_assurance /28092018Guideline_Validation_AnalyticalMethodValidation.
This work is licensed under a Creative Commons Attribution 4.0 International License.